Ambient Noise Micro‑Station is a Class 2 outdoor automatic noise monitoring system. It supports 24/7 unattended monitoring, breakpoint‑resume data transmission, over‑limit audio recording, 4G / RS485 / RJ45 communication with HJ212‑2017 protocol. IP55‑rated cabinet, suitable for construction sites, industrial parks, airports and urban functional‑area noise supervision. Optional meteorological sensors, LED display and solar‑power kit for various outdoor projects.

The Ambient Noise Micro-Station is an automatic monitoring instrument system that continuously collects, processes and analyses noise emissions at fixed monitoring points.
Its hardware consists of an all-weather outdoor microphone, a noise analysis unit, a data acquisition and transmission unit, a power supply unit, a human-machine interaction unit, an all-weather protective cabinet and supporting accessories. Optional modules include meteorological sensors, audio and video capture, and an outdoor LED display screen.
Working principle: the outdoor microphone converts sound signals into electrical signals; the noise analysis unit analyses and stores these signals; the data acquisition and transmission unit then pushes results — noise data, meteorological data, GPS position and device status — to the LED display screen for public viewing, and uploads them to the noise monitoring management cloud platform through the 4G network.
The system is designed for long-term unattended outdoor operation: automatic calibration, remote self-checking, timed restart, breakpoint resume data storage and automatic over-limit audio recording. Site managers can review real-time values, historical curves and audio clips from any browser or mobile terminal, which makes the station suitable for compliance monitoring, dispute evidence collection and long-term acoustic environment assessment.
Construction Sites: Boundary noise monitoring during earthwork, piling and demolition, supporting local construction noise limits and over-limit alarm to the site office.
Industrial Plants & Industrial Parks: Factory boundary and workshop-area noise supervision for environmental acceptance and community complaint handling.
Airports & Railway / Highway Corridors: Long-term noise monitoring along aircraft approach routes, runways, rail lines and expressway sections.
Social Life Noise Areas: Commercial streets, residential communities, night-time business zones and public squares.
Urban Acoustic Environment Functional Areas: Class 1–4 acoustic function zones, grid monitoring points, parks, schools and hospital surroundings.
Environmental Protection Authorities & Third-Party Service Providers: Municipal monitoring networks, EIA baseline surveys and outsourced operation projects.

Class 2 sound level meter performance — measurement complies with GB/T 3785.1 Class 2 and IEC 61672-2013 Class 2 requirements; filter complies with GB/T 3241 Class 2 and IEC 61260 Class 2.
All-weather outdoor microphone — resistant to wind, rain and dust; 90° incidence frequency response matches the weighting requirements of GB/T 3785.1; easy to dismantle and maintain.
Breakpoint resume data storage — cached data is re-uploaded automatically after network recovery, keeping the effective data transmission rate above 99%.
Remote self-check and automatic timing — remote diagnosis of microphone, analysis unit and communication status, plus scheduled restart and automatic sound calibration.
Over-limit automatic audio recording — 10–30 s clip length configurable; triggers on exceedance, supports active recording and remote playback for dispute evidence.
Fully automatic unattended operation — continuous real-time output, automatic data statistics (Leq, Lmax, Lmin, Lpeak, Lp) and real-time cloud upload.
Multi-channel data output — RS232/RS485 serial port, RJ45 network port and 4G wireless communication in one unit; uploads follow the HJ212-2017 protocol for direct connection to regulatory platforms.
Public transparency — supports outdoor LED display for live noise values, improving community communication on construction and industrial projects.
Optional expansion — meteorological parameters, audio and video capture, and solar power supply can be added according to project requirements.
IP55 outdoor cabinet, EMC grade A — stable operation in temperature −20 ~ 60 ℃, humidity ≤ 90% RH, atmospheric pressure 65–108 kPa.
The station is designed and validated against the following standards:
HJ 907 Technical requirements for automatic monitoring system of environmental noise
HJ 706 Technical specifications for environmental noise monitoring
HJ 1402 Technical specification for automatic monitoring of construction noise
HJ 906 Technical specifications for automatic monitoring of acoustic environment quality in functional areas
GB/T 3785.1 Electroacoustics — Sound level meters — Part 1: Specifications
IEC 61672-2013 / IEC 61260:2014
| Item | Specification |
|---|---|
| Fulfilled Standards | GB/T 3785-2010 Class 2, IEC 61672:2013 Class 2; GB/T 3241-2014 Class 2, IEC 61260:2014 Class 2 |
| Measuring Range | 24–143 dB(A); 26–143 dB(C); 31–143 dB(Z) |
| Frequency Weighting | A, C, Z |
| Time Weighting | F, S, I |
| Main Measurement Parameters | Lp, Lmax, Lmin, Lpeak, Leq |
| Protection Level | IP55; electromagnetic compatibility grade A |
| Data Storage | > 5 years (excluding audio clips); audio clip recording > 90 days |
| Audio Recording | 10–30 s configurable; automatic recording on exceedance; active recording; remote playback supported |
| Calibration Method | Automatic calibration, or standard noise source above Class 2 |
| Working Environment | −20 ~ 60 ℃, ≤ 90% RH (non-condensing), 65–108 kPa |
| Data Communication | RS232 / RS485 serial; RJ45 network port; 4G wireless; protocol HJ212-2017 |
| Power Supply | AC 220 V / 50 Hz, or customized |
| Optional Modules | Meteorological sensor set, audio & video capture, outdoor LED display, solar power kit |
| Installation | Outdoor pole-mounted, all-weather protective cabinet |
Measuring range, power supply and optional modules can be configured to match local regulation and project requirements. Please contact our team for confirmation.

Every Detail Matters. Every Product Deserves Care.
At Oceanus, quality is considered throughout the manufacturing process. From component inspection and product assembly to performance verification and final inspection, we focus on every detail to ensure our products meet the applicable quality requirements before delivery.

Pre-sales consultation: application review, installation-point planning and configuration recommendation based on local noise standards.
Product customization: parameter range, cabinet color and branding, communication protocol and optional module combinations.
Factory validation: every unit is powered-on tested, aged, acoustically calibrated with a standard sound source and inspected before packing; calibration certificate available on request.
Installation support: installation guide, wiring diagram and remote commissioning assistance; on-site guidance available for project-based orders.
Warranty & maintenance: 12-month warranty on the main unit (consumables and damage caused by misuse excluded); remote diagnosis, spare-parts supply and calibration service during the whole service life.
China purchasing service: sourcing and consolidation support for project packages.
Tell us your site environment, required sensor list, power supply condition and deployment scale. Our team will recommend suitable configuration and offer quotation.
How To Cooperate With Us?
We look forward to collaborating with more overseas gas detector and air quality monitor distributors and partners to advance the development and application of gas detection and air quality monitoring technologies .
Our Address
No. 923, Floor 9, Unit 2, NO.78, Shangding Road, Zhengzhou, China.
Phone Number
+86 15981942832
Contact:Oceanus Team
Mobile/WhatsApp:+86 15981942832
Mobile/WhatsApp:+86 13673683950
E-mail:info@china-oceanus.com
Add:No.923, Floor 9, Unit 2, NO.78, Shangding Road, Zhengzhou, China